PRO Control Solutions

SKU 400-02

microdot® PRO Control Solutions help verify that your microdot® PRO glucometer and test strips are working together properly for accurate blood glucose testing. Formulated with a known glucose concentration, they are used for routine quality checks when opening a new vial of strips, after dropping the meter, or anytime results seem inconsistent. Easy to use and designed exclusively for the microdot® PRO Blood Glucose Monitoring System.


  • Designed for use only with microdot® PRO Glucometers and microdot® PRO Test Strips

  • Helps verify that your meter, strips, and testing technique are working properly

  • Supports accurate and reliable blood glucose testing through routine quality checks

  • Contains a known amount of glucose for control testing

  • Control Range may vary by Test Strip vial, please reference range listed on each Test Strip vial when control testing

Variations

When to Use

Run a control test:

  • Before first using a new microdot® PRO glucometer

  • When opening a new vial of strips

  • At least once per week

  • If the meter is dropped

  • If meter or strips may not be working correctly

  • If test results seem inaccurate or inconsistent with symptoms

How to Use

  1. Insert a microdot® PRO Blood Glucose Test Strip into the

    meter with the blood sample reaction zone facing up.

    Make sure the test strip contact points are inserted all

    the way into the meter.

  2. Wait until the flashing blood drop and test strip icons

    appear on the lower left of the meter display screen.

  3. Press “▼” to enter the control mode. You will see “ctl”

    on the lower left of the screen. Use “▲” or “▼” to

    select control level (L1 or L3).

  4. Squeeze a drop of control solution (Level 1 or Level

    3) on the top of the control solution cap. Do not apply

    control solution to the test strip directly from the control

    vials. Note - if test strip enzyme leaches into control

    solution vials, control solutions could

    become contaminated and not work properly.

  5. Hold the meter and touch the control solution to the

    edge of the sampling end of the test strip.The control

    solution will be automatically drawn into the reaction

    area of the test strip. Wipe excess control solution

    off the cap and replace cap on control solution vial

    immediately after use.

  6. The screen will begin to count down. After 8 seconds,

    the control solution testing result will appear on the

    meter display screen. A “ctl” flag will also be attached

    when results are shown on the screen.

  7. Compare the reading on the screen to the control range

    printed on the test strip bottle (for vial test strip) or box

    (for foil-wrapped test strip). If the reading does not fall

    within the control range printed on the test strip bottle or

    box, see Control Solution Trouble Shooting on page 37 of

    the Operations Manual.

  8. Remove the used test strip either by pushing the ejection

    switch or by pulling it out with your hand and dispose in

    a proper biohazard waste container.

    Insert a new test strip to perform the next test, or press

    and hold the power button to turn off the meter.

Storage and Handling

  • Storage temperature: 39°– 86°F (4°–30°C)

  • Do not refrigerate or freeze

  • Do not use after expiration date

  • Use within 90 days / 3 months after opening

  • Record discard date after first opening

  • For use only with the microdot® PRO Meter and microdot® PRO Test Strip

Limitations

  • For in vitro diagnostic use only (outside the body)

  • Use only with the microdot® PRO Blood Glucose Monitoring System

  • Do not substitute other brands of control solution

Results and Accuracy Check

  • Results should fall within the range printed on the test strip vial or box (for Individually Wrapped Test Strips)

  • In-range results indicate the meter, strips, and technique are working correctly

  • If out of range, repeat the test before using the system for blood testing

Troubleshooting and Warnings

Out-of-range results may be caused by:

  • Incorrect testing method

  • Expired or contaminated solution

  • Damaged or outdated strips

  • Meter malfunction

  • Solution too warm or too cold

  • Not discarding first drop

  • Not shaking vial properly

Additional Info


Ordering Info

microdot® PRO Control Solutions

Product number: 400-02

Quantity: 1 Set of Control Solutions (Level 1 & Level 3)