PRO Control Solutions
microdot® PRO Control Solutions help verify that your microdot® PRO glucometer and test strips are working together properly for accurate blood glucose testing. Formulated with a known glucose concentration, they are used for routine quality checks when opening a new vial of strips, after dropping the meter, or anytime results seem inconsistent. Easy to use and designed exclusively for the microdot® PRO Blood Glucose Monitoring System.
Designed for use only with microdot® PRO Glucometers and microdot® PRO Test Strips
Helps verify that your meter, strips, and testing technique are working properly
Supports accurate and reliable blood glucose testing through routine quality checks
Contains a known amount of glucose for control testing
Control Range may vary by Test Strip vial, please reference range listed on each Test Strip vial when control testing
When to Use
Run a control test:
Before first using a new microdot® PRO glucometer
When opening a new vial of strips
At least once per week
If the meter is dropped
If meter or strips may not be working correctly
If test results seem inaccurate or inconsistent with symptoms
How to Use
Insert a microdot® PRO Blood Glucose Test Strip into the
meter with the blood sample reaction zone facing up.
Make sure the test strip contact points are inserted all
the way into the meter.
Wait until the flashing blood drop and test strip icons
appear on the lower left of the meter display screen.
Press “▼” to enter the control mode. You will see “ctl”
on the lower left of the screen. Use “▲” or “▼” to
select control level (L1 or L3).
Squeeze a drop of control solution (Level 1 or Level
3) on the top of the control solution cap. Do not apply
control solution to the test strip directly from the control
vials. Note - if test strip enzyme leaches into control
solution vials, control solutions could
become contaminated and not work properly.
Hold the meter and touch the control solution to the
edge of the sampling end of the test strip.The control
solution will be automatically drawn into the reaction
area of the test strip. Wipe excess control solution
off the cap and replace cap on control solution vial
immediately after use.
The screen will begin to count down. After 8 seconds,
the control solution testing result will appear on the
meter display screen. A “ctl” flag will also be attached
when results are shown on the screen.
Compare the reading on the screen to the control range
printed on the test strip bottle (for vial test strip) or box
(for foil-wrapped test strip). If the reading does not fall
within the control range printed on the test strip bottle or
box, see Control Solution Trouble Shooting on page 37 of
the Operations Manual.
Remove the used test strip either by pushing the ejection
switch or by pulling it out with your hand and dispose in
a proper biohazard waste container.
Insert a new test strip to perform the next test, or press
and hold the power button to turn off the meter.
Storage and Handling
Storage temperature: 39°– 86°F (4°–30°C)
Do not refrigerate or freeze
Do not use after expiration date
Use within 90 days / 3 months after opening
Record discard date after first opening
For use only with the microdot® PRO Meter and microdot® PRO Test Strip
Limitations
For in vitro diagnostic use only (outside the body)
Use only with the microdot® PRO Blood Glucose Monitoring System
Do not substitute other brands of control solution
Results and Accuracy Check
Results should fall within the range printed on the test strip vial or box (for Individually Wrapped Test Strips)
In-range results indicate the meter, strips, and technique are working correctly
If out of range, repeat the test before using the system for blood testing
Troubleshooting and Warnings
Out-of-range results may be caused by:
Incorrect testing method
Expired or contaminated solution
Damaged or outdated strips
Meter malfunction
Solution too warm or too cold
Not discarding first drop
Not shaking vial properly
Additional Info
Ordering Info
microdot® PRO Control Solutions
Product number: 400-02
Quantity: 1 Set of Control Solutions (Level 1 & Level 3)