PRO Glucometer

SKU 400-01

The microdot® PRO Blood Glucose Monitoring System combines advanced sensing technology with professional-grade accuracy in a compact, easy-to-use design. It delivers fast 8-second results, requires only a small 0.8 µL sample, features automatic coding, and stores up to 1,000 test results. Approved for Venous and Capillary whole-blood testing, the system is designed for dependable performance in multi-patient use healthcare settings.


  • Autocoded meter - no coding required!

  • FDA Approved for multi-patient use

  • 20-600mg/dL Measurement Range

  • 8 Second Test Time

  • Venous & Capillary Whole Blood Approved

  • Utilizes GDH-FAD Chemistry for less interference

  • Wide Hematocrit Range (10-70%)

  • No Interference from Over 25 Substances at Therapeutic Levels

  • Individually Wrapped Test Strips Available


Additional Info

Use only with microdot® PRO Test Strips and PRO Control Solutions

More info on the microdot® PRO Glucometer



Performing a Blood Glucose Test

To Perform a Blood Glucose Test:


  1. Insert a test strip into the device to turn on the meter. Wait until the flashing blood drop icon appears in the lower left of the screen.

  2. Gently bring the test strip to touch the drop of blood. Keep the test strips in the blood drop until the meter beeps to indicate the test strip has enough blood to test.

  3. The screen will start to count down. After 8 seconds, the glucose test result will appear on the meter display screen and be automatically stored into the meter's memory.

  4. Discard the used test strip into a biohazard waste container.

Control Solution Test Procedure

  1. Insert a microdot® PRO Blood Glucose Test Strip into the

    meter with the blood sample reaction zone facing up.

    Make sure the test strip contact points are inserted all

    the way into the meter.

  2. Wait until the flashing blood drop and test strip icons

    appear on the lower left of the meter display screen.

  3. Press “▼” to enter the control mode. You will see “ctl”

    on the lower left of the screen. Use “▲” or “▼” to

    select control level (L1 or L3).

  4. Squeeze a drop of control solution (Level 1 or Level

    3) on the top of the control solution cap. Do not apply

    control solution to the test strip directly from the control

    vials. Note - if test strip enzyme leaches into control

    solution vials, control solutions could

    become contaminated and not work properly.

  5. Hold the meter and touch the control solution to the

    edge of the sampling end of the test strip.The control

    solution will be automatically drawn into the reaction

    area of the test strip. Wipe excess control solution

    off the cap and replace cap on control solution vial

    immediately after use.

  6. The screen will begin to count down. After 8 seconds,

    the control solution testing result will appear on the

    meter display screen. A “ctl” flag will also be attached

    when results are shown on the screen.

  7. Compare the reading on the screen to the control range

    printed on the test strip bottle (for vial test strip) or box

    (for foil-wrapped test strip). If the reading does not fall

    within the control range printed on the test strip bottle or

    box, see Control Solution Trouble Shooting on page 37 of

    the Operations Manual.

  8. Remove the used test strip either by pushing the ejection

    switch or by pulling it out with your hand and dispose in

    a proper biohazard waste container.

    Insert a new test strip to perform the next test, or press

    and hold the power button to turn off the meter.

Ordering Info

microdot® PRO Glucometer Kit (includes PRO Blood Glucose Meter, High Visibility Protective Sleeve, and Carrying Case

Product number: 400-LG


microdot® PRO Glucometer

Product number: 400-01


microdot® PRO Glucometer High Visibility Protective Sleeve

Product number: 801-02


microdot® PRO Glucometer Large Carrying Case

Product number: 811-01

Meter Limitations

• Inaccurate results may occur in severely

hypotensive individuals or patients in shock.

• Inaccurate low results may occur for individuals

experiencing a hyperglycemic hyperosmolar

state, with or without ketosis.

• This meter has not been evaluated in the

critically ill.

• Not for use on neonates.

• Altitudes higher than 10,335 feet may cause

inaccurate results.

• Hematocrit range higher than 70% or lower

than 10% can cause inaccurate results.

• Severe dehydration may lead to inaccurate

blood glucose test result.

• For In vitro diagnostic only.

• For over the counter use.

• Alternative site testing should be performed

only during steady-state (when glucose is not

changing rapidly).

• For multiple-patient use.

• Alternative site measurements should never be

used for insulin dosing calculations

• Do not use plasma or serum samples.

• Do not test on arterial blood samples.